Equation Generator Instructions
The Equation Generator can be used to generate custom equations for use with TrendReader 2 software.
Basic Instructions:
- Select a product.
- Select a measurement channel type.
- For SmartReader & SmartReader Plus loggers, you have the option to take the calibration values of the logger into account by selecting the Use calibration checkbox.
- Enter the transducer output values (Xmin, Xmax).
- Enter the actual values (Ymin, Ymax).
- Click the Generate Equation button.
If there are no errors, the equation will be derived in the Equation box. You can copy and paste the equation into your TrendReader 2 software.
Note: For OWL - DC Volt, the equation will be derived for full scale only; adjustment should be done in TrendReader software. For more information regarding this topic, refer to Chapter 6 of your TrendReader manual.
Example 1: SmartReader
Consider using a transmitter with 4 - 20 mA to determine flow rate between 0 - 1000 liters per second.
- Select 4 to 20 mA
- Select Use Calibration (optional)
- Xmin = 4, Xmax = 20
- Ymin = 0, Ymax = 1000
Example 2: SmartReader Plus
Consider using a transducer with 1 - 5 volts output to monitor pressure in a vessel between 0 - 100 kPa.
- Select 0 to 5 V
- Check the Use Calibration box
- Xmin = 1, Xmax = 5
- Ymin = 0, Ymax = 100
Example 3: OWL
Consider using a transmitter with 4 - 20 mA output to determine PH in a liquid.
- Select DC Current
- Xmin = 4, Xmax = 20
- Ymin = 0, Ymax = 14